Technical Description
Technical Data
AXIOM Iconos R200
AXD3-340.620.13.01.02
9
/ 12
TV frame rate
max. 25 f/s; with CAREVISION 12,5; 8; 3 pulses/s
Brightness control
Constant image brightness through automatic gain control
TV standard
72 Hz progressive scan, independent of frequency
VIDEOMED DH
High-resolution TV camera with maintenance-free 1K CCD sensor for digital
fluoroscopy, fluororadiography and DSA
Dynamic response
max. 69 dB, optimized for DSA
TV matrix
Matrix 1024²
TV frame rate
max. 30 f/s; with CAREVISION 15; 7,5; 3 pulses/s
Brightness control
Constant image brightness through automatic gain control
TV standard
72 Hz progressive scanning, independent of frequency
Monitor
SIMOMED HM; high-line with flicker-free display 72 Hz; 44 cm or.. 54 cm screen
diagonal;
SIMOMED HC; 1024 lines, flicker-free display 72 Hz; 44 cm screen diagonal;
LCD monitor 18”; 46 cm screen diagonal;
Tube assembly
OPTITOP 150/40/80HC-100, cover with fan.
Focal spot nominal value
(ICE 60336):
0.6/1.0
Nominal power (IEC
60613):
40 kW/80 kW
Cold power:
52 kW/103 kW
Nominal voltage (IEC
60613):
150 kV
Anode angle:
12°
Anode heat storage ca-
pacity:
580 000 J (783 000 HU)
Anode rotation frequen-
cy:
Radiography: 150 Hz/ 180 Hz
Fluoroscopy: 20 to 30 Hz
Mean continuous pow-
er:
max. 450 W
Total filtration (IEC
60601-1-3):
2.5 mm Al at 80 kV
Summary of Contents for AXIOM Iconos R200
Page 2: ......
Page 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Page 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Page 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Page 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...