Examination
Digital Radiography
24
/ 52
AXD3-340.620.13.01.02
Operator Manual
Selecting the dominants
The dominants are filed in the organ programs or can be set on the integrated
generator control console.
Multiple selection of the dominants is possible.
Exception: central circular dominant.
◆
Press one of the keys:
Left dominant
Middle dominant
Right dominant
Central circular dominant
When collimating under fluoroscopy, take care that the selected left or right dom-
inant lies completely in the radiation field.
Otherwise there will be overexposure because of the inappropriate acquisition of
the measured exposure values.
The position of the measuring field/of the measuring field in relation to the object
can be checked on the LIH image by means of brief fluoroscopy. A diagram with
the position and size of the relevant measuring field is displayed briefly on the
monitor.
Warning
Please make sure when changing the selection of the dominants that the cor-
rect dominant is always selected.
If a selected left and/or right dominant becomes invalid temporarily, e.g. due to
collimation, then the central circular dominant is selected and displayed automat-
ically.
The originally selected dominant(s) will be active and displayed again when the
collimator is opened again.
Summary of Contents for AXIOM Iconos R200
Page 2: ......
Page 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Page 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Page 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Page 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...