Remote control commands
R&S
®
RTB2000
340
User Manual 1333.1611.02 ─ 11
TRIGger:A:PATTern:MODE
<PatternMode>
Disables the time limitation or sets the time comparison mode.
Parameters:
<PatternMode>
OFF | TIMeout | WIDTh
OFF
Disables the time limitation.
TIMeout
Defines how long at least the result of the state pattern condition
must be true or false.
WIDTh
Defines a time range for keeping up the true result of the pattern
condition. The range is defined using
Manual operation:
See
TRIGger:A:PATTern:WIDTh:RANGe
<PatternRange>
Selects how the time limit for the pattern condition is defined.
TRIGger:A:PATTern:WIDTh[:WIDTh]
, the range
.
Parameters:
<PatternRange>
WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers if the pattern condition remains unchanged longer than
Time - Delta
and shorter than
Time + Delta
.
OUTSide
Triggers if the pattern condition remains unchanged either
shorter than
Time - Delta
or longer than
Time + Delta
.
SHORter | LONGer
Triggers if the pattern condition changes before or after the
specified time.
Manual operation:
See
TRIGger:A:PATTern:WIDTh[:WIDTh]
<PatternWidth>
For the ranges WITHin and OUTSide, the <PatternWidth> defines the center of a
range which is defined by the limits ±<Delta>.
For the ranges SHORter and LONGer, the pattern width defines the maximum and
minimum values, respectively.
Parameters:
<PatternWidth>
Default unit: s
Manual operation:
See
Trigger