Waveform setup
R&S
®
RTB2000
65
User Manual 1333.1611.02 ─ 11
Two connector pins for compensation are at the front panel. The left pin is on ground
level. The next pin supplies a square wave signal for the adjustment.
1. Press the
[Apps Selection] key.
2. Tap "Probe Adjust".
3. Follow the instructions of the wizard. It guides you through the compensation proc-
ess.
Use the compensation trimmer of the probe to get optimum square wave response.
For details, refer to the documentation of your probe.
4.4.2
Probe settings
For passive probes, which are connected with a BNC connector, you set the probe
attenuation and the unit, and you can start an adjustment procedure for the probe. All
settings are channel-specific. When you set the attenuation, you can select a prede-
fined factor, for example "10:1", or enter a user-defined value.
If AC coupling is set, the attenuation of passive probes has no effect, and voltage is
applied to the instrument with factor 1:1. Observe the voltage limits, otherwise you can
damage the instrument.
Access: [Ch <n>] > "Probe" (scroll down). Or: short menu > "Probe"
Probes