Remote control commands
R&S
®
RTB2000
327
User Manual 1333.1611.02 ─ 11
CHANnel<m>:ARIThmetics
<TrArith>
Selects the method to build the resulting waveform from several consecutive acquisi-
tions of the signal.
Suffix:
<m>
.
1..4
The command affects all channels regardless of the indicated
channel number. The suffix can be omitted.
Parameters:
<TrArith>
OFF | ENVelope | AVERage
OFF
The data of the current acquisition is recorded according to the
decimation settings.
ENVelope
Detects the minimum and maximum values in an sample interval
over a number of acquisitions.
AVERage
Calculates the average from the data of the current acquisition
and a number of acquisitions before. The number of used
acquisitions is set with
*RST:
OFF
Manual operation:
See
ACQuire:PEAKdetect
<PeakDetect>
Enables or disables the peak detect acquisition mode.
You can use this command alternatively to
Parameters:
<PeakDetect>
AUTO | OFF
*RST:
OFF
Manual operation:
See
ACQuire:HRESolution
<HighRes>
Enables or disables the high resolution acquisition mode.
You can use this command alternatively to
Parameters:
<HighRes>
AUTO | OFF
*RST:
OFF
Manual operation:
See
Waveform setup