MEASurement<m>:DELay:SLOPe
<SignalSlope>,<ReferenceSlope>
Sets the edges to be used for delay measurement and delay to trigger measurement.
The associated waveforms are defined with
Suffix:
<m>
.
1..6
Selects the measurement place.
Parameters:
<SignalSlope>
POSitive | NEGative
Slope of first waveform ("Measure Source").
*RST:
POS
<ReferenceSlope>
POSitive | NEGative
Slope of the second waveform ("Measure Source 2"). The value
is ignored in delay to trigger measurements.
*RST:
POS
Manual operation:
See
MEASurement<m>:STATistics[:ENABle]
<StatisticEnable>
Activates or deactivates the statistical evaluation for all active measurements.
Suffix:
<m>
.
The suffix is irrelevant.
Parameters:
<StatisticEnable>
ON | OFF
*RST:
OFF
Manual operation:
See
MEASurement<m>:STATistics:RESet
Deletes the statistical results for all measurements, and starts a new statistical evalua-
tion if the acquisition is running. The waveform count is set to 0 and all measurement
values are set to NAN.
Suffix:
<m>
.
The suffix is irrelevant.
Usage:
Event
Manual operation:
See
15.7.2.2
Measurements results
You can query the statistical results using the
MEAS:STAT
commands.