Serial bus analysis
R&S
®
RTB2000
269
User Manual 1333.1611.02 ─ 11
After setting the "Compare" condition, you can enter the identifier value by setting the
state high, low, or X (do not care) for each single bit. Alternatively, you can enter a hex-
adecimal value for each half byte.
The settings are the same as for the setup of the identifier trigger, see also
If a label list with node names was loaded and applied in the bus configuration, you
can select the node name from the list instead of entering the numeric identifier. The
instrument triggers on the identifier of the selected node.
Remote command:
SEARch:PROTocol:LIN:ICONdition
SEARch:PROTocol:LIN:IDENtifier
Data condition
Settings to define the data pattern to be searched, if "Event" = "Id & Error" is selected.
After setting the "Data" length and the "Compare" condition, you can enter the data
value by setting the state high, low, or X (do not care) for each single bit. Alternatively,
you can enter a hexadecimal value for each half byte.
The settings are the same as for the setup of the data trigger, see also
Remote command:
SEARch:PROTocol:LIN:DCONdition
12.6.6
LIN label list
For general information on label lists, see
Label lists are protocol-specific. Label lists for LIN are available in CSV and PTT for-
mat.
A LIN label file contains two values for each identifier:
●
Identifier value
●
Symbolic name for the identifier
Example of a LIN PTT file
# ----------------------------------------------------------------------------
@FILE_VERSION = 1.0
@PROTOCOL_NAME = lin
# ----------------------------------------------------------------------------
# Labels for LIN protocol
# Column order: Identifier, Label
# ----------------------------------------------------------------------------
# Labels for standard addresses
0x3F,Temperature
1Ch,Left brake
20h,Right brake
LIN (option R&S
RTB-K3)