Measurements
R&S
®
RTB2000
132
User Manual 1333.1611.02 ─ 11
Figure 7-1: Results of four active measurements
Measurement errors are indicated as follows:
?
A result cannot be determined. Adjust the horizontal and vertical settings if
the instrument cannot measure.
"c" or "clipping-"
The measurement result is outside the measurement range and clipping
occurs. Adjust the vertical scale to get valid results.
n/a
No valid waveform is available, for example, if the source waveform is off.
Remote commands are described in:
●
Chapter 15.7.2.2, "Measurements results"
7.2.1.1
Statistics
In addition to the current measurement results, you can enable a statistic evaluation. It
returns the current, minimum and maximum measurement values, the average and
standard deviation, and the number of measured waveforms. The results are shown in
a separate tab below the grid. If the cursor measurement is active simultaneously, its
results are shown beside the statistic results.
Automatic measurements