Getting started
R&S
®
RTB2000
32
User Manual 1333.1611.02 ─ 11
As an alternative, you can mount several products in a rack.
4.
NOTICE!
Overheating can damage the product.
Prevent overheating as follows:
● Keep a minimum distance of 10
cm between the fan openings of the product
and any object in the vicinity.
● Do not place the product next to heat-generating equipment such as radiators
or other products.
2.1.4.2
Mounting the product in a rack
The instrument can be installed in a rack using a rack adapter kit. The order number is
given in the data sheet. The installation instructions are part of the adapter kit.
To prepare the rack
1. Observe the requirements and instructions in
2.
NOTICE!
Insufficient airflow can cause overheating and damage the product.
Design and implement an efficient ventilation concept for the rack.
To mount the R&S
RTB2000 in a rack
1. Use an adapter kit that fits the dimensions of the R&S
RTB2000 to prepare the
instrument for rack mounting. For information on the dimensions, see data sheet.
a) Order the rack adapter kit designed for the R&S
RTB2000. For the order num-
ber, see data sheet.
b) Mount the adapter kit. Follow the assembly instructions provided with the
adapter kit.
2. Push the product onto the shelf until the rack brackets fit closely to the rack.
3. Tighten all screws at the rack brackets with a tightening torque of 1.2
Nm to secure
the product at the rack.
To unmount the product from a rack
1. Loosen the screws at the rack brackets.
2. Remove the product from the rack.
3. If placing the product on a bench top again, unmount the adapter kit from the prod-
uct. Follow the instructions provided with the adapter kit.
2.1.5
Considerations for test setup
Observe safety instructions, see
Preparing for use