Getting started
R&S
®
RTB2000
34
User Manual 1333.1611.02 ─ 11
The required ratings are listed next to the AC power connector and in the data
sheet.
2.1.7
Switching on or off
The instrument is switched on or off with the power switch and the [Standby] key.
Table 2-1: Overview of power states
Status
LED
Power switch
Off
(unlighted)
[0] (off)
Standby
yellow
[I] (on)
Ready
green
[I] (on)
To switch on the product
The product is off but connected to power.
1. Set the switch on the power supply at the rear panel of the instrument to position
[I].
The power key lights up. The key is located in the bottom left corner of the front
panel.
2. Press the [Standby] key.
The instrument performs a system check and then starts the firmware.
The [Standby] key turns green and the illuminated keys on the front panel light up.
If the previous session was terminated regularly, the oscilloscope uses the last set-
tings.
Before you start measurements, be sure to comply with the warm-up phase specified
in the data sheet.
To shut down the product
The product is in the ready state.
► Press the [Standby] key.
All current settings are saved, and the software shuts down. The [Standby] key
turns yellow. The standby power supplies only the power switch circuits.
To disconnect from power
The product is in the standby state.
1.
NOTICE!
Risk of data loss. If you disconnect the product from power when it is in
the ready state, you can lose settings and data. Shut it down first.
Set the switch on the power supply to position [0].
The LED of the [Standby] key is switched off.