Remote control commands
R&S
®
RTB2000
342
User Manual 1333.1611.02 ─ 11
15.5.7
Serial bus
Triggering on the decoded data of a protocol bus is available if at least one serial proto-
col option is installed. The serial bus must be configured correctly. For detailed infor-
mation on serial protocols and their trigger conditions, see the chapter of the relevant
protocol.
15.5.8
Actions on trigger
To generate a pulse on the Aux
Out connector on trigger event, use
. You can also define the pulse length and polarity with
............................................................................................... 342
.............................................................................................. 343
.............................................................................................. 343
TRIGger:EVENt:SCRSave:DESTination
........................................................................... 343
........................................................................................... 344
............................................................................................. 344
TRIGger:EVENt:WFMSave:DESTination
.......................................................................... 344
TRIGger:EVENt[:ENABle]
<EventsEnabled>
Activates the selected actions on trigger event.
To activate the actions, use:
●
●
●
●
●
Parameters:
<EventsEnabled>
ON | OFF
*RST:
OFF
Manual operation:
See
TRIGger:EVENt:SOUNd
<SoundEnable>
If ON, the instrument generates a beep on trigger event. The acquisition is not delayed,
the sound generation runs asynchronously. The minimum time between two beeps is
1
s. If the instrument triggers faster, not all events are notified by a beep.
Parameters:
<SoundEnable>
ON | OFF
*RST:
OFF
Trigger