Status Messages
61
9.1.2
Using the XE EDP Command
The XE EDP command can be used to remotely query the
480
for the error conditions shown on the front panel.
The XE command returns two five digit numbers in the format:
xxxxx yyyyy
where
xxxxx
contains a decimal representation of any existing error conditions as described in Table 9-2.
If more than one error condition exists, the number returned is the sum of the values representing the error
conditions. For example, if the XE command returns the number 1040, this value represents the sum of an A/D
reference error (1024) and an A/D calibration checksum error (16).
The second number returned (
yyyyy
) uses the same bit assignments as shown in Table 9-2 to indicate whether the
test for the error condition was run. For example, the value
yyyyy
= 50815 represents the decimal equivalent of the
binary value 1100 0110 0111 1111. Using the bit assignments in Table 9-2, this value indicates all tests were run.
Table 9-2. Error Codes Returned on XE Command
9.2
Status Messages
Two EDP commands, P and ZZ, can be used to provide status about the indicator. These commands are described
in the following sections.
9.2.1
P EDP Command
The P EDP command returns the current displayed weight value, along with the units identifier. If the indicator is
in an underrange or overload condition, the weight value is replaced with
&&&&&&
(overload) or
::::::
(underrange).
9.2.2
ZZ EDP Command
The ZZ EDP command can be used to remotely query which annunciators are currently displayed on the indicator
front panel. The ZZ command returns the currently displayed weight and a decimal number representing the LED
annunciators currently lit. The format of the returned data is:
wwwwww uu
zzz
where
wwwwww uu
is the current displayed weight and units, and
zzz
is the annunciator status value (see Table 9-3).
If more than one annunciator is lit, ZZZ is the sum of the values representing the active annunciators.
Error
Code
Description
Binary Value
1
EEPROM Error
0000 0000 0000 0001
2
Virgin EEPROM
0000 0000 0000 0010
4
Config Parameter Checksum
0000 0000 0000 0100
8
Load Cell Checksum
0000 0000 0000 1000
16
A/D Calibration Checksum
0000 0000 0001 0000
32
Print Formats Checksum
0000 0000 0010 0000
64
XA Internal RAM Error
0000 0000 0100 0000
128
External RAM Error
0000 0000 1000 0000
256
Reserved
0000 0001 0000 0000
512
ADC Physical Error
0000 0010 0000 0000
1024
ADC Reference
0000 0100 0000 0000
2048
Count Error
0000 1000 0000 0000
4096
Reserved
0001 0000 0000 0000
8192
Display Range
0010 0000 0000 0000
16384
ADC Range
0100 0000 0000 0000
32768
Gross Limit
1000 0000 0000 0000
0x10000 - 0x80000000
Reserved
Summary of Contents for Legend 480-2A