60
480 Legend Series
9.0
Appendix
9.1
Error Messages
The
480
indicator provides a number of error messages. When an error occurs, the message is shown on the
indicator LED display.
9.1.1
Displayed Error Messages
The
480
provides a number of front panel error messages to assist in problem diagnosis. Table 9-1 lists these
messages and their meanings.
Shorting the excitation voltage shuts the excitation voltage off. The only way to restore excitation voltage is to
cycle power.
Error Message
Description
Solution
E A/D
A/D physical error
Call Rice Lake Weighing Systems Service at 800-472-6703.
EEEROM
EEPROM physical error
EVIREE
Virgin EEPROM
Use TEST menu to perform DEFLT (restore defaults) procedure, then
recalibrate load cells.
EPCKSM
Parameter checksum error
EACKSM
A/D calibration checksum error
A/D converter requires recalibration. Call Rice Lake Weighing
Systems Service.
EFCKSM
Printer format checksum error
Call Rice Lake Weighing Systems Service at 800-472-6703.
ELCKSM
Load cell calibration checksum error
Recalibrate load cells.
EIDATA
Internal RAM checksum error
Call Rice Lake Weighing Systems Service at 800-472-6703.
E REF
A/D reference error
A/D converter requires recalibration. Call Rice Lake Weighing
Systems Service.
ERROR
Internal program error
Check configuration.
Call Rice Lake Weighing Systems Service if unable to clear error by
cycling power or if error recurs.
OVERFL
Overflow error
Weight value too large to be displayed.
- - - - - -
- - - - - -
- - - - - -
Gross > overload limit
Gross value exceeds overload limit. Check configuration or signal
input level. Overload can be caused by input signal > 45 mV or
common mode voltage > 950 mV.
- - - - - -
Gross < 20d behind zero
Gross value is more than 20 divisions behind zero (OIML).
RNGERR
GRADS > 100,000
WVAL > 100,000
Only shows up in configuration mode.
EEPERR
EEPROM error
Call Rice Lake Weighing Systems for service at 800-472-6703.
HINOFF?
High offset
Zero load at powerup is more than initial zero range (INIZR) setting of
calibration zero
–
remove the extra load.
LINOFF
Low offset
Zero load at power up is less than initial zero range (INIZR) setting of
calibration zero
–
add the missing load.
NOBATT
No battery
The RTC lost time/date tracking at previous power off state due to
low battery or no battery condition. The printer, accumulator and
audit functions will fail to get time and date.
Refers to internal coin battery only, not the
rechargeable battery option.
EUCKSM
Configuration checksum
The checksum value of configuration has changed from that stored
in memory.
OIMLER
OIML parameter error
Parameter set incorrectly for use in the OIML mode.
Example: Primary units set for lb or oz.
EE-ACC
Accumulator error
Error with the accumulator such as attempting to display an
accumulated value greater than six digits.
Table 9-1.
480
Error Messages
Note
Note