Configuration Using the Front Panel (Legal for Trade)
37
3.2.4
Calibration Menu
See Section 4.0 on page 46 for Calibration procedures.
Figure 3-12. Calibration Menu
Parameter
Choices
Description
WZERO
CAL
Do not adjust this value after WSPAN has been set.
Press
to display ENTER, then press
to capture the zero value.
EDIT
Press
to display the current captured zero mV value. The value may be edited as required.
Press
to save the edited value.
WVAL
number
Press
to display and edit the test weight value.
WSPAN
CAL
Press
to display ENTER, then press
to capture the net span value.
EDIT
Press
to display the current captured net span mV value. The value may be edited as
required. Press
to save the edited value.
REZERO
ZERO
(Flashing)
Press
to remove the weight of the testing apparatus like hooks or chains used during the
calibration process.
Always use this parameter after WZERO and WSPAN have been set to recapture a new zero
value. See Section 4.1 on page 46 for more information about using this parameter.
Table 3-13. Calibration Menu Parameters
WZERO
WVAL
WSPAN
REZERO
Display and edit
zero calibration
mV value
Display and edit
span calibration
mV value
Press Enter to
remove offset from
zero and span
calibrations
Display and edit
test weight value
ENTER
Press Enter to
capture the span
net mV value
CONFIG
FORMAT
CALIBR
PROGRM
DIG IN
ALGOUT
PASWRD
USBMEM
MENU
TEST
ZERO
EDIT
CAL
000000
EDIT
CAL
Press Enter to
capture the
zero mV value
ENTER
EDIT is the current calibrated value that can be edited.
Note