![R&S CMW-KM5 Series User Manual Download Page 840](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802840.webp)
LTE SRS Measurement
R&S
®
CMW-KM5xx/-KS5xx
840
User Manual 1173.9628.02 ─ 16
The individual segments of a list mode measurement with R&S CMWS use differ-
ent connector benches. This is not allowed. All segments must use the same
bench.
Check the "Info" dialog for the relevant segment numbers.
●
101 (Firmware Error)
:
Indicates a firmware or software error. If you encounter this error for the first time,
restart the instrument.
If the error occurs again, consider the following hints:
–
Firmware errors can often be repaired by restoring the factory default settings.
To restore these settings, restart your instrument and press the "Factory
Default" softkey during startup.
–
If a software package (update) has not been properly installed this is often indi-
cated in the "Setup" dialog, section "SW/HW-Equipment > Installed Software".
–
A software update correcting the error may be available. Updates are e.g. pro-
vided in the "CMW Customer Web" on GLORIS (registration required):
.
If you get firmware errors even with the properly installed latest software version,
please send a problem report including log files to Rohde & Schwarz.
●
102 (Unidentified Error)
:
Indicates an error not covered by other reliability values. For troubleshooting
please follow the steps described for "101 (Firmware Error)".
●
103 (Parameter Error)
:
Indicates that the measurement could not be performed due to internal conflicting
parameter settings.
A good approach to localize the conflicting settings is to start with a reset or preset
or even restore the factory default settings. Then reconfigure the measurement
step by step and check when the error occurs for the first time.
If you need assistance to localize the conflicting parameter settings, please contact
Rohde & Schwarz (see
http://www.service.rohde-schwarz.com
5.5.2 General Measurement Settings
The commands valid for all LTE measurements are described here:
"General Measurement Settings"
5.5.3 SRS Measurement Commands
The commands for the "LTE SRS" measurement are divided into the groups listed
below.
Measurement Control and States
......................................................................... 841
.....................................................................................843
Power Dynamics Results (Traces)
........................................................................848
Power Dynamics Results (Single Values)
.............................................................849
Command Reference
深圳德标仪器
135-1095-0799