![R&S CMW-KM5 Series User Manual Download Page 17](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802017.webp)
LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
17
User Manual 1173.9628.02 ─ 16
Fig. 2-6: Test setup for scenario "IQ out - RF in" (example)
2.2.1.6
Test Setup for External Fading Scenarios
In order to superimpose fading on the baseband signal, you can integrate an
R&S
AMU200A into a test setup. The R&S AMU200A must be connected to the digital
I/Q interface of the R&S
CMW. At least one I/Q board must be installed at the
R&S
CMW for that purpose (option R&S CMW-B510x/-B520x).
All connections between R&S
CMW and R&S AMU200A are established via the rear
panels of the instruments.
The following figure shows a setup with two downlink paths using the first I/Q board
(DIG IQ 1 to 4).
General Description