5.
Raise or lower the sensor mounting pad to the desired position.
6.
Verify that at least one pair of access holes in the sensor mounting pad is
aligned with a pair of mounting screw holes in the Y axis arm on Cobra.
Adjust the position of the mounting pad if necessary.
7.
Screw the two M4 socket-head cap screws removed in Step 4 into the sensor
mounting pad access holes, and tighten. Use a 3 mm hex key (Allen)
wrench)
8.
Perform Steps 4 through 8 of the
Rotational Alignment
procedure described
previously in this section to verify the rotational alignment of the sensor
mounting pad.
9.
Reinstall the DRS laser sensor, as described in Section 4.
This completes the procedure for adjusting the vertical position of the sensor
mounting pad.