13. Click on the
Differential
icon or select
Differential
in the
Cursors
menu.
14. In the Output window, select
Height (Avg)
in the first measurement
drop-down box.
15. Click and drag the primary cursor boundaries, so that the cursor includes
the top of a column, as shown below.
16. Click and drag the reference cursor boundaries, so that the cursor includes a
flat area next to the column, as shown below.
-
The height of the column in the primary cursor is displayed in the
Output window next to
Height (Avg)
, and should be between 0.06 and
0.09 mm.
17. In the Output window, check the box next to the
Height (Avg)
.
-
The software automatically lists the height results in the Report View
window.
3-10
Getting Started