Section 2
Installation and Setup
This section describes how to install and set up your Cobra, install the Scan-X
software, set up the parallel port address, enable the Scan-X 3D mode, and
register Scan-X.
System Certification
All Cobra systems are inspected and certified at the factory prior to shipment
using NIST-traceable step standards. The systems are certified only for the
current mounting position of the sensor mounting pad (Z slide).
In special circumstances, it may be necessary to alter the position of the Z slide
to fit a specific application if the slide cannot be moved into the desired position
using the Focus knob. For example, a tall part may require more working
clearance thereby necessitating a higher mounting position of the Z slide.
Note:
If the mounting position of the sensor mounting pad (Z slide) is altered in
any way from the factory installed position, your Cobra should be re-certified to
ensure system accuracy. See Section 5,
Adjusting the Sensor Mounting Pad
, for
more information on raising or lowering the position of the mounting pad.
Consult your local authorized sales representative for more information on field
certification of your Cobra system.
Installation and Setup
2-1
Installation
and
Setup
Summary of Contents for Cobra DRS-300
Page 1: ......
Page 5: ...This page was left blank intentionally Cobra Laser Profile Scanner Users Guide...
Page 53: ...This page was left blank intentionally 2 32 Installation and Setup...
Page 67: ...This page was left blank intentionally 3 14 Getting Started...
Page 83: ...This page was left blank intentionally 5 6 Adjusting the Sensor Mounting Pad...
Page 87: ...This page was left blank intentionally 6 4 Troubleshooting...
Page 91: ...This page was left blank intentionally A 4 Setting the Display Properties...
Page 100: ......