4.
Insert the other two partially-threaded screws into the upper two screw
holes in the camera mounting plate and sensor. Then use a 3 mm hex key
(Allen) wrench to tighten the screws firmly (see Figure 4-6).
Never connect or disconnect the laser sensor cable from the sensor while power
is applied to Cobra. Doing so could seriously damage the sensor.
5.
Connect the laser sensor cable to the top of the sensor.
6.
Check the camera focus and alignment, as described in Appendix B.
This completes the procedure for mounting the video camera assembly and
replacement sensor.
Swapping Sensors
4-7
Swapping
Sensors
Figure 4-6. Installing the Upper Two Screws