52
7
W
Factory default
Data = 0xAA55
8
R
Type ID
0x0102
10
R
Measured value distance MSW
32-Bit
11
Measured value distance
LSW
"
22
R/W
Configuration text:
Bit-0
= 0 = QA-not active
1 = QA-active
Bit-1
= 0 = 0 - 20 mA
1 = 4 - 20 mA
Bit-2
= 0 = in case of error set
measured value to null
1 = last measured value
Bit-3
= 0 = Q3 switching output
1 = Q3 alarm function –
switches when error as in reg.
132 occurs
Bit-4
= free
Bit-5
= free
Bit-6
= free
Bit-7
= free
0x0007
112
R/W Average
0 (OFF) , 1, 2 …
100
0
113
R/W
Measurement frequency
(measurement range) MSW in
[ms]
0=Auto, 100 …
86.000.000
(24h)
32-Bit
0
114
Measurement frequency
(measurement range) LSW
"
121
W
Reset, a new sensor start is
activated
Parameters do
not change
123
R/W
Time before trigger is activated
[ms]
0, 1, ... 60 000
124
R/W
Define the type of edge to
activate trigger
0 = rising edge
1 = falling edge
2 = both edge
1