Fig. 34: In this example, the measurement is taken on cell chamber
3.
1. Cell chamber 1
2. Cell chamber 2
3. Cell chamber 3
4. Cell chamber 4
5. Measuring tip of the multimeter
Note the voltage measured on the maintenance schedule.
10.2.9 Regeneration of the Water Hardness Meter
1.
Replace the measuring cartridge with the reserve measuring
cartridge.
2.
Regenerate/soften the measuring cartridge removed.
A2012
I.
II.
III.
Fig. 35: Regenerating the measuring cartridge
1.
I: Use the regeneration kit provided. Use saturated brine for
regeneration. For example use salt from the brine storage
tank.
2.
II: Flush the measuring cartridge five times with brine. The
white part of the cartridge should move right to the other side
of the measuring cartridge.
Maintenance checks and preparation for maintenance