18.B. Installation Procedure (continued)
9. Slide the inlet tube holder onto
the vertical support rod.
7213T
A
10. Attach the fitting at the end of
the inlet tubing to the outside
port of the injector pump. For
dual-injector systems, ensure
the injector tubing goes to the
correct outer port. Refer to the
diagram located below the
pumps for details.
7017T
A
11. Snap the stainless steel tube
securely into the slot found on
the inlet tube holder.
7014T
A
12. Plug one end of the DB-15 cable
into the back of the injector
system console and the other
end to the connector labeled
Pump Connection on the back
panel of the instrument.
Note:
Use only the cable provided with the instrument. Plugging an
unauthorized cable into the connector may result in an electrical short and
will void the warranty.
Outlet
Inlet
!