22A. Specifications (continued)
UV-Vis Absorbance Module Specifications
•
Light Source: Xenon Flash Lamp
•
Detector: Photodiode
•
Spectral Range: 200–1,100nm
•
Filters Wheel Capacity: Holds up to six filters: four installed filters and two
customizable filter paddles with 260nm and 280nm preinstalled
•
Wavelengths Available: 260, 280, 450, 560, 600, 750nm
•
Photometric Measuring Range: 0–5.0 OD
•
Linear Dynamic Range: 0–3.0 OD
•
OD Accuracy: 0.05 OD ± 3%
•
OD Precision: 0.002 OD ± 3%
•
Stray Light: 0.001% at 560nm in clear-bottom, black-walled plates
Fluorescence Module Specifications
•
Light Source: wavelength matched LED
•
Detector: PiN-photodiode
•
Read Position: top reading
•
Wavelength Selection: snap-in optical kits
•
Wavelengths Available:
• UV (Ex: 365nm, Em: 410–460nm)
• Blue (Ex: 490nm, Em: 510–570nm)
• Green (Ex: 525nm, Em: 580–640nm)
• Red (Ex: 625nm, Em: 660–720nm)
•
Custom kits upon request
•
Detection Limit: 0.5fmol/200µl, or 1ppt of fluorescein in 96-well microplate
•
Linear Dynamic Range: 6 logs (assay dependent)
Injector System Specifications
•
Number of Injectors: single or dual injectors
•
Dispense Volume Range: 25–200µl in 5µl increments
•
Waste Collection Tray Volume: ~50ml
•
Dead volume: ~450µl
Shaker System Specifications
•
Linear and Orbital shaking pattern
•
Shaking Intensity: Low, Medium and High (150, 300 and 500rpm,
respectively)
•
Shaking Interval: 0.1–120 minutes
Heater System Specifications
•
Ambient 20°–45°C ± 0.75°C
•
Heating Interval: 0.1–120 minutes
Promega Corporation ·
2800 Woods Hollow Road
·
Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526
·
Phone 608-274-4330
·
Fax 608-277-2516
· www.promega.com
Part# TM311
Printed in USA.
Page 88
Revised 5/10