6. Using the provided 3/16 inch Allen wrench, tighten the four 3/4 inch Allen
head screws.
7. Using the provided 1/4 inch Allen wrench, remove all four feet from the
bottom of the GloMax
®
-Multi+ Detection System (Figure 53).
8. Reattach each of the Instrument feet and the provided spacer with the 1.5
inch Allen head screw using the provided 1/4 inch Allen wrench
(Figure 54).
9. With all four extended instrument feet attached, set the GloMax
®
-Multi+
Detection System back on its feet.
10. Manually open the instrument door and install the UV-Vis Absorbance
Module. See Section 17.C for complete instructions.
11. Power on the instrument.
Figure 53. Removing the feet from the bottom of the instrument.
12. To ensure that the instrument is detecting the newly installed UV-Vis
Absorbance Module, verify that the “Absorbance” icon appears on the
Instrument status bar of the Instrument Control screen (Figure 50).
Figure 54. Spacer and 1.5 inch Allen head screw.
7961T
A
7962T
A