4. Drop the injector tip end(s) of
the outlet tubing assembly
through the opening of the
instrument as shown.
7013T
A
5. The black plate connected to
the injector tip should be
oriented so that the screw holes
are aligned with the angle
found on the instrument.
Tighten the two black screws
with the provided Allen
wrench.
7216T
A
6. Manually open the instrument
door and hold it down. Insert
the injector tip into the injector
tip holder on the front of the
Luminescence Module.
Note:
Injector tip #1 is on the
left, and tip #2 is on the right.
7215T
A
Note:
During installation, take care to prevent damage to the injector tip.
Any nicks or deformations can adversely affect the quality of the injection
and mixing.
7. Attach the fitting at the end of
the outlet tubing assembly to
the inside port of the injector
pump. For dual-injector
systems, ensure the line to
Injector 1 goes to the inside
port of the left pump and the
line for Injector 2 goes to the
inside port of the right pump.
7022T
A
8. On the injector system console,
insert the thicker end of the
vertical support rod into the
small opening. Press down
firmly until the rod clicks in
place.
7995T