9.D. Changing the Optical Crosstalk Mask for 384-Well Plates (continued)
4. Replace the 96-well optical crosstalk mask by sliding the 384-well optical
crosstalk head mask into place on the optical head. Align the mask with the
optical head and gently guide the optical crosstalk mask onto the optical
head (Figure 23).
Figure 23. Slide the 384-well optical crosstalk mask into place on the optical head.
5. Return the injector tip holder to the optical head. Position the holder just
above the front pins on the optical head. Push the holder toward the right
to lock the holder securely to the right pin of the optical head. Then push
down and lock the holder on the left pin (Figure 24).
6. Insert the injector tip(s) into the injector tip holder and snap into place
(Figure 25).
Figure 24. Slide the 384-well optical crosstalk mask into place on the optical head.
7942T
A
7941T
A