16.H. Disconnecting from a PC
Follow the directions below to disconnect from the PC and return control to
the GloMax
®
-Multi+ Detection System touch screen.
1. Using the PC version of the software, change the Instrument Port setting to
None.
2. Touch the “Cancel” button on the External Control Enabled dialog box of
the GloMax
®
-Multi+ Detection System touch screen.
3. Verify the transfer by checking that the Instrument Status bar on the
GloMax
®
-Multi+ Detection System touch screen reads “Ready”.
16.I. USB Flash Drive
The USB Flash Drive (Cat.# E8935) included with the GloMax
®
-Multi+
Detection System allows data to be transported to an external computer. It
includes 2GB of space and has been tested to ensure compatibility with the
GloMax
®
-Multi+ Detection System.
Most USB flash drives are compatible with the GloMax
®
-Multi+ Detection
System. USB flash drives greater than 1GB may take longer than others to be
recognized. Wait a few seconds for the instrument to recognize the USB flash
drive. If the problem persists, try a different flash drive.
Note:
Results files can only be transferred from the instrument to a computer
via a USB flash drive. Use the USB port on the front of the instrument.
17.
Detection Module Installation
17.A. Luminescence Module
The Luminescence Module is required for luminescence detection. If your
instrument does not have this module, contact Promega Technical Services for
information about purchasing and installation of the Luminescence Module.
See Section 22.D for ordering information.
Due to optical and alignment complexities, the Luminescence Module may be
installed only at authorized service centers. Please contact Promega Technical
Services for details.