15.3.1.2 Preparing adapter for test milling
NOTE
Please follow directions on general cleanliness and handling.
•
The two test pins, available from your local premilled blank supplier,
are thoroughly inserted into the middle positions (2 + 5), pushed in to
the limit, and firmly clamped with the clamping screw by using a torque
wrench.
•
The DESS adapter is then fixed into the clean milling unit reception.
The screws should always be screwed in loosely at first and tightened
afterwards in the same order, 1 - 6.
•
The test pins are milled directly on the milling unit with the scripted
test program available from your local dealer. A 2mm metal tool (T2) is
required for the milling process.
15 Adapters
58 Planmeca PlanMill 60 S
User's manual