Also clean the filter of the coolant lubricant tank regularly or replace this to
guarantee the correct flow of the cooling lubricant. Ideally, clean the filter last
of all, so that any loosened dirt does not enter into the coolant lubricant tank.
Always dispose of coolant lubricant residues and contaminants in
accordance with the legal provisions and local regulations.
17.15 Condensate container
The air pressure regulator's condensate container must be checked at least
once daily and emptied if necessary.
1. Switch off the compressed air supply and make sure the air pressure
regulator is pressure-free.
2. Turn the condensate outlet of the condensate container
counterclockwise.
3. Wait until the condensate container is completely empty.
4. To close the condensate container, turn the condensate outlet clockwise
until hand-tight.
NOTE
In the case of contamination it is essential to have the compressed air supply
checked.
17.16 Periodic inspections
Inspect the machine periodically (at regular time intervals) for wear (signs of
fatigue/corrosion) of the components. Also check general machine functions,
such as safety devices.
17.17 Annual maintenance tasks
NOTE
The annual maintenance is performed by an authorised service technician.
Visually inspect the milling unit's earthing connections every 6 months.
We recommend that milling units in multi-shift operation be subjected
to semiannual maintenance and milling units in single-shift operation be
subjected to annual maintenance of the milling unit by a Planmeca approved
service technician.
All mechanical components and machine parts are subject to increased
(natural) wear.
We recommend performing annual maintenance after approx. 12 months,
replacing the customary wear and tear parts, in order to guarantee seamless
and long-term functionality.
17 Cleaning and maintenance
User's manual
Planmeca PlanMill 60 S 85