Trilogy 202
clinical manual
74
Indicator
Description
Leak
This indicator displays the total leak (non-returned flow), between
the unit outlet and the patient, averaged over the previous
breath. This only appears in detailed view when the Passive
Circuit is selected.
Measured Settings Panel
All of the indicators that may appear on the Measured Settings panel
(available only in Detailed view), are described in the following table.
Indicator
Description
PIP
Peak Inspiratory Pressure displays the maximum pressure
delivered to the patient during the previous breath.
I:E Ratio
Displays a comparison of the time spent in inspiration to the time
spent in expiration during the previous breath.
Peak Flow
Displays the maximum inspiratory flow delivered to the patient
during the previous breath in l/min BTPS.
MAP
Displays the Mean Airway Pressure, which is the weighted
average of pressure in the patient’s airway over 6 breaths.
FiO
2
Displays the FiO
2
setpoint.
MinVent
Minute ventilation displays the amount of air delivered to the
patient over the last minute in l/min BTPS.
Measured Settings Panel