Chapter 5
Viewing and Changing Settings
85
Low Inspiratory Pressure
This setting configures the Low Inspiratory Pressure alarm. It is only
user-settable in CV, AC, and SIMV modes. It cannot be set lower than
PEEP + 2 cmH
2
O or higher than the High Inspiratory Pressure. For
passive circuits, you can increase or decrease the Low Inspiratory
Pressure from 6 to 40 cmH
2
O in increments of 1 cmH
2
O. For active
circuits, you can increase or decrease the setting from 2 to 40 cmH
2
O
in increments of 1 cmH
2
O. For pressure modes, this alarm is not user-
settable.
High Inspiratory Pressure
This setting enables or disables the High Inspiratory Pressure
alarm. It is only user-settable in CV, AC, and SIMV modes. The High
Inspiratory Pressure cannot be set lower than the Low Inspiratory
Pressure. You can increase or decrease the High Inspiratory Pressure
from 10 to 80 cmH
2
O in increments of 1 cmH
2
O. For pressure modes,
this alarm is not user-settable.
Additional Settings Specific to Therapy Modes
The Settings and Alarms menu also contains many additional
settings specific to the various therapy modes. The specific settings
for each therapy mode are described below.
Continuous Positive Airway Pressure (CPAP) Mode
In addition to the general settings described in the previous section
of this manual, you can also set the following settings in CPAP mode.
1. CPAP
You can increase or decrease the CPAP pressure setting from 4
to 20 cmH
2
O in increments of 1.
Note:
If the CPAP pressure is
set to 4 (the minimum setting),
the Ramp Length setting will be
unavailable.
Summary of Contents for Trilogy 202
Page 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Page 3: ......
Page 23: ...Trilogy 202 user manual 14 ...
Page 59: ...Trilogy 202 user manual 50 ...
Page 75: ...Trilogy 202 user manual 66 ...
Page 145: ...Trilogy 202 clinical manual 136 ...
Page 169: ...Trilogy 202 clinical manual 160 ...
Page 177: ...Trilogy 202 clinical manual 168 ...