Trilogy 202
clinical manual
100
Updating Prescriptions Using the SD Card
With the Trilogy 202, you can update the patient’s prescription
using the SD Card. This feature lets you update a single prescription
or both prescriptions if the dual prescription feature is enabled. The
prescription update can occur either when the ventilator is off or on.
1. Insert an SD Card with a valid prescription into the device. One
of the following screens appears on the display (depending on
whether the device is on or off):
WARNING
When you change the device
prescription, alarms and other
settings using the SD card,
Trilogy 202
requires that the
caregiver review and verify the
changes prior to the changes
being used by the device.
The caregiver or health care
professional is responsible to
ensure that the prescription
settings are correct and
compatible with the patient
after using this feature. Installing
the wrong prescription for a
particular patient may result
in improper therapy, lack of
appropriate safety monitoring,
and risk of death or injury to the
patient.
Ventilator Off Prescription
Change Screen
Ventilator On Prescription
Change Screen