Chapter 10
System Checkout Procedures
137
This chapter details the test procedures that should be performed by
the clinician prior to connecting the device to the patient. Test the
Active PAP, Active Flow, and Passive circuit types if you want to do a
complete checkout on the device. The tests should be performed as
described in order to verify proper operation of the device. Some of
the procedures in this chapter require you to change settings on the
device. If you are not familiar with the procedures for accomplishing
this, please refer to Chapter 4,
Ventilator Setup
and Chapter 5,
Viewing
and Changing Settings
.
Tools Required
Universal Porting Block
•
Active PAP Exhalation Device
•
Active Flow Exhalation Device with Flow Sensor
•
Trilogy Universal Active PAP Tube Adapter
•
Passive Exhalation Device
•
Test Lung
•
High Pressure O
•
2
Hose
Oxygen Monitor
•
Small Flat Head Screwdriver
•
Trilogy 202
clinical manual
10. System Checkout Procedures
Note:
The actual circuit
configuration to be used on
the patient should be used to
perform the system checkout
procedure.
WARNING
If you notice any unexplained
changes in the performance
of the device, if it is making
unusual sounds, if the device
or detachable battery are
dropped, if water is spilled into
the enclosure, or if the enclosure
is cracked or broken, discontinue
use and contact Philips
Respironics or an authorized
service center for service.
Summary of Contents for Trilogy 202
Page 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Page 3: ......
Page 23: ...Trilogy 202 user manual 14 ...
Page 59: ...Trilogy 202 user manual 50 ...
Page 75: ...Trilogy 202 user manual 66 ...
Page 145: ...Trilogy 202 clinical manual 136 ...
Page 169: ...Trilogy 202 clinical manual 160 ...
Page 177: ...Trilogy 202 clinical manual 168 ...