Chapter 3
Modes, Features, and Alarms
29
During the time window, if patient effort is detected, either a
Spontaneous or an Assist breath is delivered for the first effort. If the
last breath delivered was Mandatory, then a Spontaneous breath is
delivered for the first patient effort of the window. If the last breath
delivered was an Assist or Spontaneous breath, then an Assist breath
is delivered for the first patient effort of the window. Assist breaths
deliver the prescribed Tidal Volume during inhalation and the PEEP
setting during exhalation.
During the time window, if patient effort continues to be detected
(after the first patient triggered breath), Spontaneous breaths are
delivered for the remainder of the window. Spontaneous breaths
deliver the prescribed Pressure Support setting above PEEP during
inhalation and PEEP during exhalation. The figure below provides
example breath patterns in SIMV mode.
12 seconds
12 seconds
12 seconds
12 seconds
2
3
3
1
1
1
Tidal Volume = 500 mL
Breath Rate = 5 BPM
Inspiratory Time = 3.0 seconds
Pressure Support = 10 cm H
2
O
PEEP= 6 cm H
2
O
= Pressure Supported
Breath
= Mandatory Breath
= Assist Breath
= Mandatory Window
= Spontaneous Window
1
2
5
10
15
20
25
30
35
40
45
6
10
20
26
Enter SIMV mode
M
M
M
M
M
S
S
S
PRESSURE
TIME
2
SIMV Mode
Summary of Contents for Trilogy 202
Page 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Page 3: ......
Page 23: ...Trilogy 202 user manual 14 ...
Page 59: ...Trilogy 202 user manual 50 ...
Page 75: ...Trilogy 202 user manual 66 ...
Page 145: ...Trilogy 202 clinical manual 136 ...
Page 169: ...Trilogy 202 clinical manual 160 ...
Page 177: ...Trilogy 202 clinical manual 168 ...