Chapter 10
System Checkout Procedures
143
The Circuit Disconnect alarm condition appears on
•
the screen, highlighted in red
D. Reconnect Test Lung
Reconnect the test lung to the circuit.
E. Verify Reset
Wait at least 40 seconds and verify the following auto-reset
conditions:
The High Priority audible indicator has stopped
•
sounding
The red light on the
•
Alarm Indicator/Audio Pause
button has stopped flashing
F. Restore Ventilator Settings
Modify the ventilator settings and change the following values
shown below in Table 10.
Setting
Value
Circuit Disconnect
Off
5. Verify the High Inspiratory Pressure Alarm
This procedure verifies that the High Inspiratory Pressure alarm
is working properly. It assumes that you have attached the test
lung, verified ventilator settings, and turned on ventilator power as
described in the Initial Setup section.
A. Change Ventilator Settings
Modify the ventilator settings and change the following values
shown below in Table 11.
Setting
Value
Mode
CV
Tidal Volume
500 ml
FiO
2
21%
Breath Rate
12 BPM
Inspiratory Time
1.0 seconds
Table 10 - Restore Ventilator
Settings
Table 11 - Ventilator Settings
Note
: If this alarm is not reset
within 3 occurrences, the alarm
is elevated to High Priority, and
the High Priority Indicators
occur.
Summary of Contents for Trilogy 202
Page 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Page 3: ......
Page 23: ...Trilogy 202 user manual 14 ...
Page 59: ...Trilogy 202 user manual 50 ...
Page 75: ...Trilogy 202 user manual 66 ...
Page 145: ...Trilogy 202 clinical manual 136 ...
Page 169: ...Trilogy 202 clinical manual 160 ...
Page 177: ...Trilogy 202 clinical manual 168 ...