Chapter 10
System Checkout Procedures
147
Verify Blending
A. Turn on the ventilator.
B. Verify the set level of FiO
2
is satisfied using an external O
2
monitor.
Verify alarm
A. Shut off or disconnect the source of high pressure O
2
to the
ventilator.
B. Wait 1 minute and verify the following alarm signals:
The Priority audible indicator sounds.
•
A red light flashes on the Alarm Indicator/Audio
•
Pause Button.
The following alarm conditions appear on the screen,
•
highlighted in red:
- Low Oxygen Flow
- Low Oxygen Inlet Pressure
Verify Reset
A. Reconnect or turn on the source of high pressure O
2
to the
ventilator.
B. Wait 1 minute and verify the following:
The set level of FiO
•
2
is satisfied using an external O
2
monitor.
The High Priority audible indicator has stopped
•
sounding.
The red light on the Alarm Indicator/Audio Pause
•
button has stopped flashing.