Chapter 11
Technical Specifications
153
Control Accuracy
Parameter
Range
Accuracy
IPAP
4 to 50
1
cmH
2
O
2
Greater of 2 cmH
2
O or 8% of setting
EPAP
0 to 25 cmH
2
O for Active Circuits
4 to 25 cmH
2
O for Passive Circuits
Greater of 2 cmH
2
O or 8% of setting
CPAP
4 to 20 cmH
2
O
Greater of 2 cmH
2
O or 8% of setting
PEEP
0 to 25 cmH
2
O for Active Circuits
4 to 25 cmH
2
O for Passive Circuits
Greater of 2 cmH
2
O or 8% of setting
4
Pressure
4 to 50 cmH
2
O
Greater of 2 cmH
2
O or 8% of setting
Pressure Support
0 to 30 cmH
2
O
3
Greater of 2 cmH
2
O or 8% of setting
4
Tidal Volume
50 to 2000 ml
5
Greater of 15ml or 10% of setting (Active Circuits)
Greater of 15ml or 15% of setting (Passive Circuits)
Breath Rate
0 to 60 BPM for AC mode
1 to 60 BPM for all other modes
Greater of ± 1 BPM or ±10% of the setting
Timed Inspiration
0.3 to 5.0 seconds
± .1 second
Rise Time
1 to 6
6
±2 cmH
2
O (the device will increase to a pressure
of .67* (IPAP - EPAP) ±2 cmH
2
O @ the set rise time
multiplied by 100 ms for pressure supports less than
or equal to 25.)
Ramp Start Pressure
0 to 25 cmH
2
O for Active Circuits
4 to 25 cmH
2
O for Passive Circuits
4 to 19 cmH
2
O in CPAP mode
8% of s 2% Full Scale
Ramp Length
Off, 5 to 45 minutes
±2 minutes
Flex
Off, 1 to 3
7
N/A
Flow Trigger Sensitivity
1 to 9 l/min
N/A
Flow Cycle
10 to 90%
N/A
Apnea Rate
4 to 60 BPM
Greater of 1 BPM or 10% of setting
FiO2 Output
21% to 100%
21% to 50% is ±3%
50% to 95% is ±5%
100% is -5%
O2 Input Pressure Rating
40 to 87 PSI
N/A
Specifications listed are based on using a standard circuit consisting of 1.8 meter tubing- REF 622038, one of the following exhalation
devices: Whisper Swivel II - REF 332113; Active PAP Exhalation Device - REF 1053716; Active Flow Exhalation Device REF 1049503 or
Active Exhalation Device – REF 1065659 and Trilogy Proximal Flow Sensor- REF 1050408 (if required).
1
Limited to 25 cmH
2
O when using the Bi-Flex feature in S mode.
2
Pressure units may be cm H
2
O, hPa, or mBar depending on device setup.
3
The difference between the Inspiratory Pressure and the Expiratory Pressure must never be more than 30 cmH
2
O.
4
Pressure Support and PEEP not to exceed 50 cmH
2
O.
5
All flows and volumes are measured at BTPS conditions.
6
The range of values correspond to tenths of seconds (e.g., a setting of 4 indicates a Rise Time of 0.4 seconds).
7
Flex is not available when AVAPS is active. Flex is not available with Active Circuits.
Summary of Contents for Trilogy 202
Page 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Page 3: ......
Page 23: ...Trilogy 202 user manual 14 ...
Page 59: ...Trilogy 202 user manual 50 ...
Page 75: ...Trilogy 202 user manual 66 ...
Page 145: ...Trilogy 202 clinical manual 136 ...
Page 169: ...Trilogy 202 clinical manual 160 ...
Page 177: ...Trilogy 202 clinical manual 168 ...