Chapter 2
System Description
15
This chapter describes the front and rear panel device controls and
features.
Front Panel Features
The front panel contains the control buttons, visual indicators, and
display screen.
Buttons
The following buttons are included on the front panel of the device.
1.
Start/Stop Button
This button turns the airflow on or off, starting or stopping
therapy.
2.
Alarm Indicator and Audio Pause Button
This button serves two purposes: it temporarily silences
the audible portion of an alarm, and it also acts as an alarm
indicator. When silencing an alarm, if the cause of the alarm is
not corrected, the alarm sounds again after two minutes. Each
time the button is pressed, the alarm silence period resets to two
minutes. See Chapter 6 for more information.
3.
Up/Down Button
This button allows you to navigate the display menu and edit
device settings.
1
2
4
5
3
Front Panel Controls and
Display Screen
Note:
When you start therapy,
the display backlight and the
backlights on the buttons turn
on, the red and yellow alarm
LEDs turn on momentarily, and
an audible indicator sounds
to indicate that therapy has
started. The Startup screen
appears on the display.
WARNING
To make sure the device is operating
properly at start-up, always verify
that the audible tone sounds and
the alarm LEDs light red and then
yellow momentarily. Contact Philips
Respironics or an authorized service
center for service if these indications
do not occur at start-up.
Trilogy 202
user manual
2. System Description
Summary of Contents for Trilogy 202
Page 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Page 3: ......
Page 23: ...Trilogy 202 user manual 14 ...
Page 59: ...Trilogy 202 user manual 50 ...
Page 75: ...Trilogy 202 user manual 66 ...
Page 145: ...Trilogy 202 clinical manual 136 ...
Page 169: ...Trilogy 202 clinical manual 160 ...
Page 177: ...Trilogy 202 clinical manual 168 ...