Trilogy 202
clinical manual
142
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
The High Priority audible indicator has stopped
•
sounding
The red light on the
•
Alarm Indicator/Audio Pause
button has stopped flashing
E. Restore Ventilator Settings
Modify the ventilator settings and change the following value
shown in Table 8.
Setting
Value
Low Vte/Low Vti
Off
4. Verify Circuit Disconnect Alarm
This procedure verifies that the Circuit Disconnect alarm is working
properly. It assumes that you have attached the test lung, verified
ventilator settings, and turned on ventilator power as described in
the Initial Setup section.
A. Change Circuit Disconnect Ventilator Setting
Modify the Circuit Disconnect ventilator setting to match the
value shown below in Table 9.
Setting
Value
Circuit Disconnect
10 seconds
B. Disconnect Test Lung
Disconnect the test lung from the circuit.
C. Verify the Alarm
Wait approximately 10 seconds and verify the following alarm
signals:
The High Priority Audible Indicator sounds
•
A red light flashes on the
•
Alarm Indicator/Audio
Pause button
Table 8 - Restore Ventilator
Settings
Table 9 - Ventilator Settings
Note:
The Low Inspiratory or
Low Expiratory Pressure Alarm
may also be detected.
Summary of Contents for Trilogy 202
Page 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Page 3: ......
Page 23: ...Trilogy 202 user manual 14 ...
Page 59: ...Trilogy 202 user manual 50 ...
Page 75: ...Trilogy 202 user manual 66 ...
Page 145: ...Trilogy 202 clinical manual 136 ...
Page 169: ...Trilogy 202 clinical manual 160 ...
Page 177: ...Trilogy 202 clinical manual 168 ...