background image

 

IFU1911203 - HCSRE Issue 4.doc 

© Semmco LPS Ltd 2015 – This document is not to be copied or reproduced without the permission of Semmco LPS Ltd 

 

HEAD 15 CSRE SET CARRY CASE 

 

K02 Regenerative Device 
For Short Term Use 
 
Instruction Manual 

 
 
 
 
 
 
 

 
CONTENTS 

 

 

IMPORTANT NOTICE 

 

INTENDED USE 

 

PRECAUTIONS 

 

PRIOR TO USE 

 

OPENING DEVICE   

 

DONNING DEVICE  

 

USAGE PICTOGRAMS 

 

USE 

 

AFTER USE 

 

STORAGE AND USAGE CONDITIONS 

 

RETURN CONDITIONS 

 

ORDERING INFORMATION 

 

PRODUCT IDENTIFICATION 

 

TECHNICAL DATA 

 

EU DECLARATION 

 

IMPORTANT NOTICE 

 
1.  The equipment should only be used for the purposes specified in this 

manual. 

2.  All individuals involved in the use of the equipment must read and 

understand the contents of this manual and have been trained by a 
Semmco LPS Ltd certificated trainer. 

3.  All individuals involved with the training of the end user or 

maintenance of this equipment must be authorised and certificated by 
Semmco LPS Ltd. 

4.  The warranties made by Semmco LPS Ltd with the regard to this 

product are void if the equipment is not used and serviced in 
accordance with the instructions in this manual. 

5.  The warranties and conditions of sale of Semmco LPS Ltd are not to 

be altered by the above recommendations. 

 

INTENDED USE 

 

HEAD 15 CSRE is a short duration, oxygen breathing device, which is 
intended for a single use for Cell Snatch Rescue, in areas of: 

 

smoke 

 

toxic contamination and 

 

oxygen deficiency 

HEAD 15 CSRE is supplied in a Carry Case with a 15 minute duration. 
The apparatus, described in these Instructions for Use, has been designed 
and tested in accordance with EN 13794, EN 145 and EN 1146 where 
applicable, to demonstrate compliance with the essential safety 
requirements with PPE Regulation (EU) 2016/425. Intrinsic safety as per 
SIRA certificate Sira 04Y2057X. 
 
 
Type examination performance by: SGS United Kingdom Ltd, Unit 202B, 
Worle Parkway, Weston-super-Mare, BS22 6WA 
(Notified Body Number 0120) 
 
 

PRECAUTIONS 
 

 

The carrying case is anti-static and is therefore suitable for use in 
explosive atmospheres. 

 

Due to the potential fire risk it is important to prevent the ingress of 
flammable substances such as petrol, grease or solvents into the 
device before and during operation. 

 

If the device is damaged there is a risk that flammable substances 
may be ignited if they come into contact with the oxygen producing 
chemical. 

 

Do not remove the device from the case until the equipment is 
required for use. Doing so may reduce the protection afforded by the 
device. 

 
 
 
 
 
 

 
 

PRIOR TO USE

 

Check case to ensure:

  

 

The case is not damaged. 

 

The security seal is still intact. 

 

The unit is within the expiry date. 

 

The vacuum pack is intact and the Indicator is not pink in colour. 

 

The equipment should not be used if these conditions are not 
met.

 

 

OPENING DEVICE 

 

HEAD 15 CSRE CARRYING CASE 

1.  Open carrying bag by pulling webbing tag. 
2.  The lid will fall away to reveal vacuum pack. 
3.  Unfold tear strip, hold securely and rip open using red tear tab. 
4.  Pull firmly on the breathing protection device to remove from 

vacuum pack. This action automatically activates the oxygen 
generation into the breathing bag, commences the alarm self test 
procedure and removes the two red sealing bungs from the 
device. If these remain attached remove manually. 

 

DONNING HEAD 15 CSRE 

1.  Unfold the hood, place fingers inside neck seal, inhale and pull 

onto head. 

2.  Hold front port and adjust position of inner mask to ensure correct 

fit, comfort and field of vision. 

3.  Exhale into mask and commence normal breathing. 

NOTE: 

 

The action of removing the breathing device from the vacuum 
bag automatically activates oxygen generation, the alarm self 
test and removes the two red sealing bungs from the device. 
The LED and alarm will sound for a few seconds to indicate it 
is functioning correctly. In the unlikely event the alarm self 
check is not activated, HEAD 15 CSRE remains safe to use 
however no reliance should be placed on the alarm timer. The 
two red rubber bungs are connected inside the bag via a red 
lanyard, if for any reason the bungs stay attached to the 
device when it has been removed from the bag they should be 
removed from the device manually before donning starts. 

 

Facial hair in the sealing area may cause leaks and can 
seriously reduce the protection provided by the apparatus. 

 

The correct positioning of the inner mask is essential to 
provide a face seal and prevent ingress of atmospheric 
contaminants. 

 

Always follow training procedures. The hood must be 
properly sealed before starting to use the equipment. 

 

USAGE PICTOGRAMS 
 
HEAD 15 CSRE CC 

 

USE 

 

 

Once  activated  and  the  correct  function  checked,  the  red  LED’s 
and  audible  warning  will  operate  intermittently  when  two  thirds 
duration of the operating time has elapsed. This will change to the 
constant mode when the stated duration is reached. This warning 
will continue for a further period of approximately 5 minutes.

 

 

When the  warning  device  activates the constant mode, HEAD 15 
CSRE

 

should be removed as this indicates that oxygen generation 

may  be  ending  and  that  the  device  may  no  longer  be  providing 
protection.

 

IMPORTANT NOTES FOR USERS 

 

Immediately  the  device  is  donned  and  a  seal  obtained 
activities  should  commence  in  a  calm  manner.  Do  not 
rush. 

 

Maintain  a  steady  work  rate  to  ensure  regular  breathing. 
Rapid breathing uses more oxygen. 

 

During  use  the  starter  and  oxygen

 

generators  will 

increase  in  temperature  and  the  air  provided  by  this 
device  will  be  warm  and  dry  and  slight  odours  or  tastes 
may  be  detected.  These  are  normal  conditions  and 
indicate the device is functioning correctly. 

 
AFTER USE 

 

 

Remove the hood immediately in a safe area, or when the warning 
device indicates oxygen production may have ceased.

 

 

The device can be returned to Semmco LPS Ltd for reclamation or 
rework, see section on ‘RETURN CONDITIONS’.

 

 

- REMOVE UNIT 

Reviews: