Trilogy 202
user manual
34
Flow Pattern Types
Two flow patterns are available in Volume Control ventilation therapy
modes:
Square
•
Ramp
•
Square
With a square wave pattern, airflow is generally constant throughout
inspiration of the breath.
FLOW
TIME
Ramp
With a ramp flow pattern, the airflow starts high and decreases
throughout inspiration of the breath.
50% of
Peak Flow
FLOW
TIME
Square Flow Pattern
Ramp Waveform Pattern