Chapter 10
System Checkout Procedures
141
E. Restore Ventilator Settings
Modify the ventilator settings and change the following value
shown in Table 5.
Setting
Value
High Vte/High Vti
Off
3. Verify the Low Tidal Volume Alarm
This procedure verifies that the Low Tidal Volume alarm is working
properly. For Passive and Active Flow circuits, this will verify the Low
Vte alarm. For Active PAP circuits, this will verify the Low Vti alarm. It
assumes that you have attached the test lung, verified the ventilator
settings, and turned on ventilator power as described in the Initial
Setup section.
A. Change Alarm Ventilator Setting
Modify the Low Tidal Volume alarm setting to match the one
shown below in Table 6.
Setting
Value
Low Vte/Low Vti
500 ml
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
The High Priority audible indicator sounds
•
A red light flashes on the
•
Alarm Indicator/Audio
Pause button
The Low Tidal Volume alarm condition appears on the
•
screen, highlighted in red
C. Modify Ventilator Alarm Settings
Modify the Low Tidal Volume alarm setting to match the one
shown below in Table 7.
Setting
Value
Low Vte/Low Vti
50 ml
Table 5 - Restore Ventilator
Settings
Table 6 - Low Vte/Low Vti
Alarm Setting
Table 7 - Modify Low Vte/Low
Vti Alarm Setting
Summary of Contents for Trilogy 202
Page 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Page 3: ......
Page 23: ...Trilogy 202 user manual 14 ...
Page 59: ...Trilogy 202 user manual 50 ...
Page 75: ...Trilogy 202 user manual 66 ...
Page 145: ...Trilogy 202 clinical manual 136 ...
Page 169: ...Trilogy 202 clinical manual 160 ...
Page 177: ...Trilogy 202 clinical manual 168 ...