Chapter 7
Cleaning and Maintenance
123
Cleaning the Ventilator
The ventilator’s exterior surface and the exterior of the detachable
battery pack (if using) should be cleaned before and after each
patient use, and more often if needed.
1. Unplug the device and clean the front panel and exterior of the
enclosure as needed using a clean cloth dampened with any of
the following cleaning agents:
– Water
– Soapy water or a mild detergent
– Hydrogen Peroxide (3%)
– Isopropyl Alcohol (91%)
– 10% bleach solution (10% bleach, 90% water)
2. Do not allow any liquid to drip into the ventilator case or
detachable battery pack. After cleaning, use a soft, dry cloth to
remove any residual cleaner. Use extra care when cleaning the
display. Abrasive cleaners can scratch the display.
3. Allow the device to dry completely before plugging in the power
cord.
WARNING
To avoid electrical shock, always
unplug the power cord from the
wall outlet before cleaning the
ventilator.
CAUTION
Do not immerse the device or
allow any liquid to enter the
enclosure or the inlet filter.
CAUTION
Do not use harsh detergents,
abrasive cleaners, or brushes to
clean the ventilator system. Use
only the cleaning agents and
methods described in this manual.
CAUTION
Do not attempt to sterilize or
autoclave the ventilator.
Trilogy 202
clinical manual
7. Cleaning and Maintenance
Summary of Contents for Trilogy 202
Page 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Page 3: ......
Page 23: ...Trilogy 202 user manual 14 ...
Page 59: ...Trilogy 202 user manual 50 ...
Page 75: ...Trilogy 202 user manual 66 ...
Page 145: ...Trilogy 202 clinical manual 136 ...
Page 169: ...Trilogy 202 clinical manual 160 ...
Page 177: ...Trilogy 202 clinical manual 168 ...