Chapter 10
System Checkout Procedures
145
6. Verify the Low Inspiratory Pressure Alarm
This procedure verifies that the Low Inspiratory Pressure alarm
is working properly. It assumes that you have attached the test
lung, verified ventilator settings, and turned on ventilator power as
described in the Initial Setup section.
A. Change Ventilator Settings
Modify the ventilator settings and change the following values
shown below in Table 13.
Setting
Value
Mode
CV
Tidal Volume
500 ml
FiO
2
21%
Breath Rate
12 BPM
Inspiratory Time
1.0 seconds
Flow Pattern
Ramp
PEEP
4 cmH
2
O
Sigh
Off
Circuit Disconnect
Off
Low Inspiratory Pressure
40 cmH
2
O
High Inspiratory Pressure
60 cmH
2
O
Apnea
Off
All other alarms
Off
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
The High Priority audible indicator sounds
•
A red light flashes on the
•
Alarm Indicator/Audio
Pause button
The Low Inspiratory Pressure alarm condition
•
appears on the screen, highlighted in red
Table 13 - Ventilator Settings