66
Preparing Your Laboratory
Table 2-8.
External Cooling Specifications
PC Configuration and Printer
Only the currently shipping PerkinElmer supplied PCs will be installed free of
charge. Only for these PCs is operation guaranteed. There will be no support for
any other PC, as no compliance testing has been done to operate our software
efficiently. For details on computer requirements see
Installation Preparation
Check for Damage
•
The customer must check the shipping box for damage. If any is found, the
carrier must be notified immediately.
Unpacking
•
The spectrometer must be unpacked and brought into the laboratory at least
one day prior to installation to allow acclimation. The customer must be
warned about the weight of the instrument and informed that at least three and
preferably four adults are necessary to lift the instrument.
Water Temperature
20-40 °C
Water Pressure
min. 350 kPa (3.5 bar, 50.8 psig)
max. 400 kPa (4.0 bar, 58 psig)
Water Flow Rate
2.5 L/min
pH of the Water
Between 6.5 and 7.5
Hardness of the Water
Not greater than 14 °d or 2.5 mol/L
Caution
If the water is too cold or flowing too fast, moisture may condense on the
graphite contacts, on the furnace windows or inside the water cooled
furnace power supply resulting in malfunction and severe damage.
Moisture on the contacts causes rapid attrition of the graphite surface,
while moisture on the furnace windows leads to attenuation of the energy.
Summary of Contents for PINAACLE 900 Series
Page 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Page 2: ......
Page 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Page 52: ...36 Safety Practices...
Page 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Page 84: ...68 Preparing Your Laboratory...
Page 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Page 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Page 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Page 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Page 210: ...194 Preparing the System...
Page 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Page 354: ...338 Maintenance...
Page 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Page 389: ......