168
Preparing the System
Table 5-2.
Required Pressures for the Furnace Gases
•
Set the outlet gauge pressure for each gas to a value just below the
maximum value (see Pmax column).
•
Never
set the outlet gauge pressure to a value exceeding the maximum
pressure (see Pmax column).
Installing the Lamps
The spectrometer has an eight position motor controlled lamp selector, and is
designed to hold eight PerkinElmer Lumina hollow cathode lamps (HCLs) or two
electrodeless discharge lamps (EDLs) and six PerkinElmer Lumina hollow cathode
lamps (HCLs). With the exception of the PinAAcle 900Z model, the spectrometer
also comes with a background corrector deuterium lamp (D
2
).
For information on lamp maintenance and cleaning see
.
Warning: Toxic Gases
If you use nitrogen as the inert gas, toxic cyanogens gas (CN)
2
will be
produced when the furnace temperature is above 2300 °C.
Make sure that the main laboratory fume ventilation system removes all
gases and fumes emitted by the furnace.
!
Warning
Summary of Contents for PINAACLE 900 Series
Page 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Page 2: ......
Page 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Page 52: ...36 Safety Practices...
Page 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Page 84: ...68 Preparing Your Laboratory...
Page 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Page 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Page 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Page 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Page 210: ...194 Preparing the System...
Page 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Page 354: ...338 Maintenance...
Page 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Page 389: ......