Maintenance
2. Unscrew the retaining ring (see figure below) and pull out the burner o-ring.
Figure 6-15.
Exploded view of the burner chamber assembly
Item
Description
Item
Description
1
Burner head
5
Stainless steel nebulizer (is
shown here but you can also
have a high sensitivity plastic
nebulizer attached here)
2
Burner head adjustment tool
6
End cap assembly
3
Spray chamber
7
Retaining ring
4
Flow spoiler
8
Burner head o-ring
1
2
7
4
8
3
6
5