227
Drain System Maintenance
Emptying the Drain System
Figure 6-18.
Drain bottle attached
1. Extinguish the flame, allow the burner head to cool, shut down the gases at
source, and bleed the gas supply lines.
2. Switch off the spectrometer.
3. Flush the drain system. See
Cleaning (or Flushing) the Drain System on page
Item
Description
1
Retainer cap
2
Drain trap assembly
3
Drain bottle
1
2
3