253
Removing and Replacing the Deuterium Arc Background Corrector Lamp
2. The access to the D
2
lamp compartment cover is located on the right side of the
instrument. Use a Phillips screwdriver to remove the screw that holds the cover
in place. Remove the cover.
Figure 6-26.
Accessing the D
2
lamp compartment cover
3. Locate the deuterium lamp assembly in the instrument. Loosen the
thumbscrew that holds the assembly in the instrument.
Item
Description
1
D2 lamp compartment cover
2
Phillips screw
Caution
As you slide out the D
2
lamp assembly take care not to damage the cable.
2
1
Summary of Contents for PINAACLE 900 Series
Page 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Page 2: ......
Page 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Page 52: ...36 Safety Practices...
Page 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Page 84: ...68 Preparing Your Laboratory...
Page 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Page 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Page 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Page 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Page 210: ...194 Preparing the System...
Page 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Page 354: ...338 Maintenance...
Page 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Page 389: ......